Thermo Icap Manual



  1. Thermo Icap Manual Download
  2. Thermo Icap 7400
  3. Thermo Icap Pro


The use of Inductively Coupled Plasma source (ICP) and Atomic Absorption (AA) are the accepted and most powerful techniques for the analysis of and quantification of trace elements in both solid and liquid samples. Applications range from routine environmental analyses to the materials industry, geological applications to clinical research and from the food industry to the semiconductor industry. The iCAP 6300-Duo can simultaneously detect 66 Elements with detection limits less than 1μg/L. The plasma is easier to ignite and can operate with a much wider range of sample types. The Duo instrument provides flexibility, with extreme detection limits achievable in axial mode and reduced interference in Radial mode.

Thermo Icap ManualThermo Icap Manual

Thermo Scientific iTEVA Software Manual vii Figure 1-1. ITEVA login.1-5 Figure 1-2.

Thermo Icap ManualThermo icap manual

Thermo Icap Manual Download

Manual

Thermo Icap 7400

  1. Key benefits. The Thermo Scientific™ iCAP™ 7000 Series ICP-OES provides the lowest cost of analysis through Thermo Fisher Scientific. ICAP Q Operating Manual (P/N 1288090, Revision B) i. Contacting Us. There are several ways to contact Thermo Fisher Scientific.
  2. The Thermo Scientific iCAP Q ICP-MS offers dramatically improved levels of performance, reliability, ease of use and flexibility compared to its predecessor, the Thermo Scientific XSERIES 2. The iCAP Q is the result of a series of technological and useability improvements derived from Thermo Scientific's wide of range of analytical instruments.
  3. The Thermo Scientific iCAP 7600 ICP-OES is the ideal solution for the most demanding analytical challenges. The instrument has the highest throughput, sensitivity and detection limits. Productivity is increased by the integrated sample loop which efficiently delivers the sample to the plasma.

Thermo Icap Pro

Technical Specification

Spectrometer

Echelle type

52.91 grooves/mm ruled grating

383 mm effective focal length

9.5 0UV fused silica cross dispersion prism

Wavelength range

166 – 847 nm

Spectral bandpass

7 pm at 200 nm

Detector

High performance CID86 Chip

RF source

27.12 MHz solid state

750 – 1500 watts output power

(Duo restricted to 1350 watts)

Sample Pump

3 – channel, 12 roller peristaltic.

Speed 0 -125 rpm

Plasma gas

Fixed 12 L/min, argon

Nebulizer gas

Pressure control, form 0 - 0.4 Mpd

Auxiliary gas

4 fixed flows, 0, 0.5, 1.0 and 1.5 L/min

Standard sampling kit

Concentric glass nebulizer

Glass cyclone spray chamber

Semi – demountable torch

1.5 mm bore quartz injector (radial version)

2 mm bore quartz injector (Duo version)

Dimensions

840 W* 750 D *590 H